Exergen: FDA Study Reveals Misleading Labeling by NCIT Manufacturers Regarding Accuracy

Since the start of the pandemic, non-contact infrared thermometers (NCITs) have been widely used as a screening method for fever detection in healthcare and public settings, as fever is a leading sign of Covid. Now, a new FDA published study1 demonstrates their failure to reliably detect fevers when used on adults in accordance with the NCIT instructions, and fail to be within the accuracy specifications as advertised in manufacturers’ literature, instructions for use, and other labeling. The study suggests that because of their high probability for producing false negative readings close to the CDC fever threshold, NCITs are an unreliable temperature screening tool. FDA’s 510(k) premarket notification database shows that more than 20 NCITs have been cleared by the FDA in the past three years.

The study, which was approved by the FDA Institutional Review Board (IRB), was undertaken to determine the accuracy of NCITs and evaluate their adherence to FDA labeling requirements. It tested six different commercially available NCIT models among a large sample of 1,113 adult subjects, 93% between 18 and 30 years old and 7% older than 30.   A total of 60 NCITs were tested, with 10 units per model. The temperatures were compared with those taken with a reference oral thermometer in the monitor mode. The mean difference between the reference thermometer and the NCIT measurements for under- and over-reporting varied widely by model. Depending on the specific device, 48% to 88% of the individual temperature measurements fell outside the labeled accuracy stated by the manufacturers. Notably, these false readings were obtained in an highly controlled environment that carefully managed all possible variables.

“This important study underscores what the medical community has known all along: NCITs are not accurate and device manufacturers are being irresponsible when accuracy matters most. It also shines light on NCIT manufacturers’ lack of adherence to FDA specifications for accuracy in labeling,” said Francesco Pompei, Ph.D., CEO of Exergen Corporation. “Covid has changed our world and we can no longer tolerate the rampant false temperature readings we get from NCITs.  As the FDA states in the study: ‘To successfully screen and track people with elevated temperature, it is essential that accurate temperature measurements are made, and that the thermometer outputs are correctly interpreted.’  We agree with the FDA’s assessment.  For everyone’s safety, NCIT manufacturers should be held responsible by the FDA so that their proven inaccuracy can be clearly seen and addressed.”

en_USEnglish